SYS 1000 ci
Efficient Performance, Powerful Results!
Superior Performance for Quality Results: High reliability and precision, user-friendliness Compact Design: Space-saving design that fits well in laboratory setups without compromising on performance. Integration Capabilities: Integration with CLIA supports efficient data handling and enhances the functionality. Continuous sample rack loading, Easy integration with immunoassays Robust and Efficient: High Quality Integration, 1,200 tests per hour with ISE, Automatic removal of reagent bubbles. Effective Optical System: Concave grating system for accurate reading. It uses Halogen Tungsten lamp (up to 2000 hours) Effortless Sample Handling: Continuous sample loading upto 180 samples at a time. Can load upto 30 samples in STAT priority mode.
Technical Specifications
| Technical Specifications: | |
| System Function | |
| Throughput: 1,000 photometric tests per hour, up to 1,200 tests per hour with ISE | |
| Measuring principles: Absorbance photometry, turbidimetry. Methodology: End-point, Fixed-time, Kinetic, ISE (optional) On-board tests: Up to 66 (60 photometric tests, 3 ISEs and 3 serum indexes) | |
| Sample Handling | |
| SDM: 180 samples by racks, sample continuous loading, STAT in priority, start in 1 minute | |
| Integrated decapper: Remove tube caps at a speed of 300 tubes/hour, UV disinfection | |
| Sample volume: 1-25 µL, with an increment of 0.1 µL. Sample probe: Liquid level detection and tracking, clot detection and collision protection. Ultrasonic washing | |
| Reagent Handling | |
| Reagent positions: 218 positions with cooling system (2-8°C) | |
| Reagent volume: 10-200 µL, with an increment of 0.5 µL Reagent probe: Liquid level detection, collision protection, bubble detection and automatic removing | |
| Reaction System | |
| Reagent volume: 80-280 µL | |
| Reaction cuvettes: 206 reusable cuvettes with an optical path of 5 mm | |
| Reaction temperature: 37-0.3°C, with ±0.1°C fluctuation | |
| Reaction cuvette wash: 8-phase cuvette wash | |
| Optical System, | |
| Light transmission mode: Reversed optics with holographic concave flat-field gratings | |
| Wavelength: 16 wavelengths (340nm, 380nm, 412nm, 450nm, 480nm, 505nm, 546nm, 570nm, 605nm, 630nm, 660nm, 700nm, 740nm, 770nm, 800nm, 850nm) | |
| Absorption range: 0-3.5Abs | |
| Control and Calibration | |
| Calibration modes: Linear (one-point, two point and multi-point). Logit-Log 4P, Logit-Log SP, Spline, Exponential, Polynomial, Parabola | |
| Quality control: Westgard multi-rule. Twin plot and L-J graphic rule | |
| Operation Unit | |
| Operating system: Windows 10 (64-bit) | |
| Interface: TCP/IP network port and USB port | |
| Network connection: Unidirectional and bidirectional LIS | |
| Environmental requirements | |
| Power Supply: 110V: 110V/115V, 60Hz 220V: 220V-240V~, 50Hz 220V/230V~, 60Hz | |
| Temperature: 15-30°C | |
| Humidity: 35-85% | |
| Water consumption: 35L/H | |
| Altitude: -400-4000m (106-61.6kPa). (For areas above 2,000 m, extemal air pump is required) | |
| Dimension: 1,720 mm*900 mm*1,230 mm (W*D*H) | |
| Weight: 450kg |