With more than 2500 installations and still counting “Photometer 5010” is referred to as “Gold Standard” for semi-automation in Indian In-vitro diagnostics. Photometer 5010 is commonly known for its everlasting performance because of its robust hardware, unique features and user friendly software. Photometer 5010 is equipped with infrared bubble detector which allows the test to be performed at minimum sample volume negating any chances of bubbling in sample. Photometer 5010 is also equipped with most advanced optical system ensuring accurate results. Its most advanced “German Hellma” flow cell which is regarded as the best in terms of quality ensuring precise light control. Photometer 5010 is equipped with LED (light emitting diode) which has thirteen times more life span as compared to a normal quartz halogen lamp with least power consumption while maintaining consistent intensity of light. Moreover, Photometer 5010 is equipped with stainless steel peristaltic pump with PEP tubing’s ensuring minimal maintenance and maximum performance.

Photometer 5010 is also equipped with most advanced temperature control (peltier system) which allows the option to perform the assay at 250C , 300C and 370C . Photometer 5010 is a product which specifically caters to the needs and wants of its customers by ensuring forever performance.

Technical Specifications

Light Source LED
Wavelength Range 340 nm - 800 nm; Automatic - 9 position filter wheel; 6 Standard filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm, 623 nm; 3 Position Free for Optional Filter of Choice; Half bandwidth: ± 2 nm
Photometric Range 0.0 to 2.5 A
Flowcell Micro Flow Cell Measuring Volume - 32 µL; Interchangeable with Normal Standard Cuvettes (Macro or Semi - Micro, Disposable or Special Optical Glass); Optical Path: 10 mm
Temperature Control Internal Peltier Element, Temperature Variable; Pre Adjusted to 25˚C, 30˚C and 37˚C; Equilibration time for Aspirated reaction mixture to reach 37˚C from ambient temperature : 15 sec
Aspiration System Built in Peristaltic Pump driven by Stepper Motor; Programmable Aspiration Volume controlled by infrared light bubble detector; Aspiration Minimum 250 µL, typically 500 µL upto 2000 µL; Separate setting of Aspirate Volume and Wash Volume
Operator Interface Touch Screen for direct functions and Alphanumeric Inputs
Data Presentation Graphic Display: white characters or symbols, blue backlight; Dimension 240x128 dots
Integrated Printer Thermal Printer; 24 Characters per line
Languages English; German
Memory General Operating Software contained in Flash Memory; Flash memory can be updated by PC (data file downloaded from Internet, floppy disk or CD ROM); Capacity for 250 pre-programmed tests; Upto 230 tests can be stored in RAM. Import of data by touchscreen or PC; Upto 50 Non-Linear Calibration Curves with max 20 sets of points can be stored; Data logging upto 2970 results of max 99 samples can be managed
Signal Port Serial port for connection to an external printer and/or PC
Measurement Procedures Absorbance, End Points with factor, Standard or multiple standards, with or without reagent blank and/or sample blank; Kinetics with standard, factor or multiple standards, with or without blank; Fixed time with standard, factor or multiple standards, with or without reagent blank Turbidimetry; Single, Double and triple determinations; Curve fitting for non-linear standard curves
Quality Control Upto 50 methods can be controlled with two control seras
Measuring Time Kinetic: Variable from 3 - 28 deltas; Time per delta 4 - 255 sec in 1 sec steps; Fixed time : Variable from 0 - 65535 sec in 1 sec steps
Delay Time Programmable from 0 - 65535 sec in 1 sec steps2
Dimension 1. 33 cm (L) X 34 cm (W) X 18 cm (H)
Weight 1. 6.0 kg